UV-A Intensity: What’s Special About 5,000 µW?

  •  

Most UV-A lamps on the market are set to provide 5,000 µW/cm2 at a distance of 15 in / 38 cm. What’s so special about those numbers?

By David Geis, Product Manager

Physics of Light

Light intensity follows an inverse-square law. Say there’s a bright lamp that’s one meter away from you. If you move out to 2 meters, the light from the lamp won’t be half as bright, it’ll be one quarter as bright. Going farther out to 3 meters drops it down to 1/9th the original intensity. It works the same way as you move closer too. The same lamp will be 4 times as bright 50 cm away, and 16 times as bright at only 25 cm. What this means for us is that the distance to the lamp is just as important as the amount of light it puts out.

To compare the intensity of UV-A lamps in fluorescent inspection, our industry has set a standard distance of 15 in / 38 cm as a benchmark. Using a standard distance, the relative performance of different lamps can be compared directly.

 

Human Vision

UV-A irradiance is required to make penetrant and magnetic particle materials fluoresce (that’s what makes them visible). The more UV-A available, the brighter the fluorescence will be. But there are limits. ASTM E2297 and E1316 define UV-A light as wavelengths between 320-400 nm and define visible light as wavelengths between 400-760 nm. But in nature, the light spectrum is continuous. There is no natural difference between what we call UV-A and what we call visible light – those are artificial divisions. Human vision, in the same way, is continuous. There’s no hard-line cutoff of what the eye can perceive, and light perception changes as we age. Younger eyes can perceive light down to 390 nm (even though we technically define that as UV-A), and if there is no other light source, even shorter wavelengths down to 380 nm can be seen as deep violet light.

In the dark environment of an inspection booth, human vision can perceive some UV-A light. Even though fluorescence increases with increasing UV-A, there are diminishing returns as far as what you can see. Too much UV-A can overpower a fluorescent indication – this is commonly called “veiling glare”.

 

UV-A Intensity Studies

Studies on the effect of UV-A intensity on fluorescence have demonstrated that the dyes and pigments used in penetrants and magnetic particle materials can degrade with long exposure to very intense irradiation. Exposure to high levels of UV-A over time will decrease the amount of fluorescent light produced. This effect is commonly described as “fluorescent fade”.

To guard against both veiling glare and fluorescent fade, maximum UV-A intensities have been established. The details vary with industry standard, OEM, Prime, and application. Many authorities permit UV-A intensities up to 10,000 µW/cm2, but exposure time must be limited to prevent fluorescent fade. The maximum intensity of 5,000 µW/cm2 according to ISO 3059 is commonly accepted as safe for extended exposures.

Learn 3 Facts About LED Optics and The Impact of Ultraviolet Light in this blog article.

 

Conclusion

We have a standard distance of 15 in / 38 cm thanks to the way that light works. And we have a maximum intensity of 5,000 µW/cm2 thanks to how human vision and fluorescence work. It’s easy to use those two figures together: 5,000 µW/cm2 at 15 in / 38 cm. With so many different light sources available, that combination doesn’t work in all cases.

The concerns leading to maximum intensity are about the intensity at the examination surface regardless of lamp distance. While a hand-held lamp may be commonly used at the standard 15 in / 38 cm distance, a torch or flashlight will be used much closer. And overhead stationary light fixtures are used at much greater distances than standard. In all cases, the UV-A intensity at the examination surface needs to be controlled. A torch that delivers 5,000 µW/cm2 at a distance of 5 in / 13 cm is very useful for close examination of bores and tight spaces. But at the standard 15 in / 38 cm, the intensity would be less than 560 µW/cm2. In the same way, an overhead lamp fixture delivering 5,000 µW/cm2 while mounted 2.5 ft / 0.76 m above the inspection surface would be nearly 10,000 µW/cm2 at the standard distance.

In these examples, the torch and the overhead fixture have widely different intensities at the standard distance. But when they are used as designed and intended, they are perfectly acceptable. That’s because ISO 3059 and other standards set the maximum intensity limit of 5,000 µW/cm2 at the examination surface, not at a standard distance of 15 in / 38 cm.

 

Solution

The ST700 was designed to meet inspection lamp specifications while maximizing the area available to perform examinations. The ST700 is a stationary lamp fixture intended to be mounted at a distance from examination surfaces, giving the inspector ample room to work. To deliver enough UV-A to perform examinations while mounted up to 36 in / 0.9 m away, the ST700 is set for 7,000 µW/cm2 at the standard distance of 15 in / 38 cm. When working to Rolls-Royce and Airbus specs, that translates to a minimum working distance of 24 in / 61 cm.

See how the ST700 stationary UV inspection lamp can eliminate or prevent the most common UV illumination challenges fluorescent NDT testing.

 

 

Published August 14, 2018

JOIN OUR MAILING LIST TO RECEIVE THE LATEST NDT INSIGHTS AND ARTICLES FROM MAGNAFLUX

Want to stay up on the latest NDT insights and articles from Magnaflux? 

Subscribe to Magnaflux updates above to get fresh news delivered to your inbox once a week.

Please wait while we gather your results.

Related Blog Entries

Why Does ASTM E3022 Require a Filter for LED UV Lamps?

Learn about the function and importance of UV-a pass filters, and why ASTM E3022 requires all LED UV-A lamps to have them.

Read More...

Understanding the Magnaflux Shelf Life Statement for NDT Chemicals

Understanding the Magnaflux Shelf Life Statement for NDT Chemicals

Answering popular questions regarding shelf life for magnetic particle and liquid penetrant products used in nondestructive testing

Read More...

Understanding Safety Data Sheets

Understanding Safety Data Sheets: Do You Know What’s in the Chemicals You Handle?

We highlight the 5 major sections to understand on a Safety Data Sheets (SDS) in accordance with US OHSHA Hazcom 2012 GHS regulations

Read More...

How to Select the Right UV-A Lamp for Your NDT Needs [Checklist]

4 things to consider when looking for a new LED UV light for fluorescent penetrant inspection or magnetic particle testing.

Read More...

Shedding Light on ASTM E2297 & E3022: A Guide to NDT UV Lamp Standards

An overview of two of the most common standards for NDT UV light sources, ASTM E2297 and ASTM E3022.

Read More...

Learn how to achieve a high intensity beam of light with UV LEDs for NDT inspection

NRTL: What Does It Mean and Why Does It Matter?

NRTL stands for Nationally Recognized Testing Laboratory, but what does that mean to you?

Read More...

What You Need to Start-up In-house Testing [Checklist]

What You Need to Start-up In-house Testing [Checklist]

Step-by-step process listing everything from the equipment needed in magnetic particle testing and liquid penetrant testing to preparation for NDT audits

Read More...

Handheld LED UV Lamps

About the EV6000 Handheld LED UV Lamp

Check out how Magnaflux’s EV6000 stacks against other leading UV-A portable handheld lamps in the market

Read More...

EV6000 Hand-held UV Lamp Demo and Q&A [Webinar]

Watch and listen to a 20-minute EV6000 demonstration by two outstanding NDT experts who afterwards host a 20-minute Q&A for participants to ask their own LED UV questions.

Read More...

ST700 Hood Mount is a freely-adjustable mounting frame

New Mounting System for Overhead UV-A Lamps

The ST700 Hood Mount is a freely-adjustable mounting frame to give operators the option to position their overhead UV-A Lamp to optimize inspection ease

Read More...

Magnaflux ST700 Overhead NDT Inspection Lamp

Stationary LED UV Lamp for NDT Inspections

See how the ST700 UV inspection lamp​ can eliminate the most common UV illumination challenges fluorescent NDT testing

Read More...

Learn how to achieve a high intensity beam of light with UV LEDs for NDT inspection

3 Facts About LED Optics and The Impact of Ultraviolet Light

Learn how to achieve a high intensity beam of light with UV LEDs for NDT inspection

Read More...

4 Monthly Checks Required for all LED Inspection-Grade Lamps

4 Monthly Checks Required for all LED Inspection-Grade Lamps

User-check requirements for LED UV-A lamps, and how to plot the beam of an LED UV-A lamp for Airbus AITM6-1001 Issue 11

Read More...

Complying with Airbus AITM requirements – How to check UV stability

Complying with Airbus AITM Requirements – How to Check UV Stability

Learn how to comply with the new requirements of the latest revision of Airbus AITM 6-1001

Read More...

Defining IP Ratings for Electronics Used in Mag Particle Testing & Liquid Penetrant Testing

Defining IP Ratings for Electronics Used in Mag Particle Testing & Liquid Penetrant Testing

In this article, we break down the ANSI/IEC 60529 Ingress Protection Code with a chart to show what's essential for NDT equipment

Read More...

NDT UV LED Lamps

Frequently Asked Questions About NDT UV LED Lamps

Frequently Asked Questions About NDT UV LED Lamps

Read More...

Nadcap AC7114 Rev J

How to Comply with NADCAP AC7114 Requirements for LED UV Lamps

An overview of Rev J changes to Nadcap AC7114 Non-Destructive Testing audit checklists for Penetrant and Magnetic Particle

Read More...

Airplane Engine NDT

How to Improve Fluorescent NDT Process Control with LED UV Lighting [Case Study]

See how an aerospace OEM manufacturer increased NDT inspection reliability & control with wide-beam, overhead LED UV-A lamps

Read More...

ST700 High Volume Inspections

How to Increase High-Volume NDT Inspection Speed and Reliability with LED UV Lighting [Case Study]

See how an automotive parts manufacturer decreased inspection time and improved reliability with wide-beam, overhead UV lamps

Read More...

Dye Penetrant Testing on Pipe

NDT UV Lamps “Stabilize” vs “Warm Up”, What’s the Difference?

Understanding the difference between NDT mercury-vapor lamp “warm-up” time and LED black light “stabilization” time

Read More...

New High-Intensity, Dual-Light LED UV Lamp for NDT Pros

Learn about how the EV6500 UV lamp for nondestructive testing improves visibility & versatility while minimizing time to inspect parts

Read More...

ST700 Overhead Inspection UV Lamp

New Stationary LED Inspection UV Lamp for NDT Pros

Learn how this new UV lamp for nondestructive testing solves the biggest challenges in fluorescent NDT UV illumination.

Read More...

UV-A Intensity: What’s Special About 5,000 µW?

UV-A Intensity: What’s Special About 5,000 µW?

Most UV-A lamps on the market are set to provide 5,000 µW/cm2 at a distance of 15 in / 38 cm. What’s so special about those numbers?

Read More...

Why Do I Get Over 2 Foot-Candle / 20 Lux Visible Light from My LED UV-A Lamp?

Why Do I Get Over 2 Foot-Candle / 20 Lux Visible Light from My LED UV-A Lamp?

A quick history of the visible light emissions requirement, where it is today, and an overview of UV meters

Read More...

JOIN OUR MAILING LIST TO RECEIVE THE LATEST NDT INSIGHTS AND ARTICLES FROM MAGNAFLUX

Magnaflux

155 Harlem Avenue
Glenview, IL 60025, USA
Telephone: +1 847-657-5300
Contact Magnaflux Customer Service

Select Your Country North America Mexico Brazil China Europe India New Zealand, Australia, Japan, Southeast Asia
© 2024 Magnaflux - All Rights Reserved.
top